Chapter 8 Maintenance
8-5
Disinfections
CAUTION: Use only recommended disinfectants on system surfaces.
A disinfectant qualified by the FDA 510(k) process is recommended. The following disinfectants are
recommended because of both their biological eectiveness (as qualied through the FDA 510(k) process)
and their chemical compatibility with Samsung Medison ultrasound products.
Solutions Country Type Active ingredient FDA 510(k)
Cidex
USA Liquid Gluteraldehyde K934434
Cidex Plus
USA Liquid Gluteraldehyde K923744
1. Turn o the system and disconnect the system power cord from the wall outlet.
2. Mix the disinfection solution compatible with your system according to label instructions for solution
strength.
3. Wipe the system surfaces with the disinfectant solution, following the disinfectant label instructions for
wipe durations, solution strength, and disinfectant contact duration.
4. Air dry or towel dry with a sterile cloth according to the instructions on the disinfectant label.
Accuracy Check
NOTE: The user must ensure that safety inspections are performed every 2 years according to the
requirements of safety standard EN 60601-1. Only trained persons are allowed to perform these
safety inspections.
The product’s maintenance status may aect the measurements obtained using the product. The product
should be maintained in an optimal state to ensure reliable measurements.
To ensure optimal operation of the product, perform an accuracy check every year. The equations and table
related to measurement accuracy are included in “Chapter 5 Measurements” in this manual.