SAMSUNG MEDISON DIAGNOSTIC ULTRASOUND SYSTEMSONOACE R5User ManualVolume 1

Strona 2

Precautions For UseHow to Use Your ManualThis manual addresses the reader who is familiar with ultrasound techniques. Only medical doctors or person

Strona 3

4-14User Manual M Mode Screen M Line Use the trackball on the control panel to move left or right. The M line indicates the relative position of the

Strona 4

4-15Chapter 4 Diagnosis Modes M Mode SoftmenuFollowing items are displayed on Softmenu in M mode. Select the item and adjust the value with the Softm

Strona 5

4-16User ManualColor Doppler ModeThis mode displays the colored blood ow pattern of the ROI (Region of Interest) within the 2D image. It is appropria

Strona 6

4-17Chapter 4 Diagnosis Modes C Mode Screen ROI BoxROI stands for Region Of Interest. In Color Doppler Mode, the ROI box outlines the area of the 2D

Strona 7

4-18User Manual InvertInverts the color bar in the opposite direction. Inverting the color bar also inverts the colors displayed on the image. Color

Strona 8 - Safety Requirements

4-19Chapter 4 Diagnosis ModesXTag Pos: It is available when Color Tag is on. Use the Menu dial-button to select the position of tags, from between 0

Strona 9 - Declarations:

4-20User ManualPower Doppler ModeThis mode displays the color intensity of blood ow within the ROI in the 2D image. It is appropriate for examining t

Strona 10 - Precautions For Use

4-21Chapter 4 Diagnosis Modes ROI BoxROI stands for Region Of Interest. In Power Doppler Mode, ROI indicates the area of the 2D image where color (b

Strona 11

4-22User ManualPW Spectral Doppler ModePW stands for Pulse Wave. PW Spectral Doppler Mode gives information on the speed of blood ow at a specic sit

Strona 12 - Revision History

4-23Chapter 4 Diagnosis Modes PW Spectral Doppler Mode ScreenOn the Spectral Doppler Mode image, the horizontal axis represents time, and the vertica

Strona 13 - Table of Contents – Volume 1

DANGER: Describes precautions necessary to prevent user hazards of great urgency. Ignoring a DANGER warning will risk life-threatening injury.WARNING:

Strona 14 - Chapter 3 Starting Diagnosis

4-24User ManualXMoving Sample Volume If you move the Sample Volume in the D Only state, the system will calculate the PRF value and the Phantom Gate

Strona 15 - Table of Contents

4-25Chapter 4 Diagnosis Modes PW Spectral Doppler Mode MenuFollowing items are displayed on menu when you press the Menu dial-button in PW Spectral D

Strona 16

4-26User ManualXEdit: Adjust a post map.XReturn: Return to the previous step of the menu.NOTE: For information on other menu items, please refer to

Strona 17 - Chapter 1

4-27Chapter 4 Diagnosis Modes Combined Modes2D/C/PW ModeColor Doppler Mode and PW Spectral Doppler Mode are displayed simultaneously.Press the PW bu

Strona 18

4-28User ManualAs each active image mode has its own set of menu items and buttons, you should press the Set button to change the active image mode.

Strona 19 - Purpose of Use

4-29Chapter 4 Diagnosis Modes Dual ModeThe multi-image viewing function is only available when the images and conditions on both sides are identical

Strona 20 - Safety Signs

4-30User Manual 3D Mode3DThe 3D Mode displays tissues or objects in the human body as three-dimensional images instead of two-dimensional images.The

Strona 21 - Chapter 1 Safety

4-31Chapter 4 Diagnosis ModesAcquiring a 3D Image The following explains how to acquire a 3D image. 1. Press the 3D key on the keyboard.2. The ROI

Strona 22

4-32User Manual3D View3D images can be seen through the 3D View application.[Figure 4.9 3D View] Viewing a 3D Image XSelect the menus using the Track

Strona 23 - Electrical Safety

4-33Chapter 4 Diagnosis Modes 3D Image Optimization Threshold (Low, High) AdjustmentAdjust the slide bar of the threshold using the Trackball and th

Strona 24

Revision HistoryVERSION DATE NOTEv1.02.02-00 2014.10.15 Initial ReleaseSystem Upgrades and Manual Set UpdatesSamsung Medison Ultrasound is committed

Strona 25

4-34User Manual[Figure 4.10 Softmenu - Rotation Mode]Tips!Other ways to rotate the image1. The ROT slide bar at the bottom of the screen can be adjus

Strona 26

4-35Chapter 4 Diagnosis Modes MagnicationTo magnify or reduce the image size, enter Magnication mode by pressing the Magnication button, and move

Strona 27 - Peripherals

4-36User Manual Multi planar Press Multi planar, and 3D multiplanar screen appears. Shows volume data in cross-section images of Coronal Section, Sag

Strona 28

4-37Chapter 4 Diagnosis ModesUse X/Y/Z slide on the screen to rotate reference plane.AYYZXXYZZXCB[Figure 4.17 The standard coordinates systems of Mul

Strona 30

Chapter 5Measurements And CalculationsUser ManualMeasurement Accuracy ... 5-3Causes of Measurement Errors ...

Strona 32

5-3Chapter 5 Measurements And Calculations Measurement AccuracyMeasurement values can vary, depending on the nature of the ultrasound, the body’s re

Strona 33 - Precautions on Ramps

5-4User Manual Ultrasound VelocityXThe velocity of ultrasound used during measurement is usually 1,540 m/s on average.XThe velocity of ultrasound ma

Strona 34 - Safety Note

5-5Chapter 5 Measurements And CalculationsOptimization of Measurement Accuracy 2D ModeXResolution is in proportion to the frequency of the probe.XP

Strona 35 - Safety Note for Monitor

13Table of Contents Table of Contents – Volume 1Chapter 1 SafetyPurpose of Use ...

Strona 36 - Biological Safety

5-6User Manual Color/Power Doppler ModeXA protocol is not specied for images in Color Doppler Mode or Power Doppler Mode. Therefore, the same limit

Strona 37 - Direct Controls

5-7Chapter 5 Measurements And Calculations 2D ModeMeasurementsSystem Tolerance:(Whichever is greater)TestMethodologyAccuracy(1)Based onRange(2)Axial

Strona 38 - Indirect Controls

5-8User Manual Basic MeasurementsPress the Caliper button on the control panel. NOTE: Take basic measurements of distance and area regardless of the

Strona 39 - Output Display Features

5-9Chapter 5 Measurements And Calculations Basic Measurement OperationsThe following is the information on common button operations for basic measure

Strona 40 - Thermal Index (TI) Display

5-10User Manual2. Use the trackball and the Set button on the control panel to specify both end points of the measurement area. XPlace the cursor a

Strona 41

5-11Chapter 5 Measurements And Calculations 2D Hip Joint This is a basic measurement that is available in all diagnosis modes. You can measure the an

Strona 42 - 2D-mode Controls

5-12User Manual M DistanceThis is a basic measurement that is available in M Mode only. You can specify two points in an M image and measure the dista

Strona 43 - Color and Power Controls

5-13Chapter 5 Measurements And CalculationsThe equations used for D Velocity measurement are as follows: Doppler Velocity (A/B)This is a basic measur

Strona 44 - M-mode and Doppler Controls

5-14User Manual3. When the measurement is nished, its result is shown on the screen.XdT: Change in TimeXVTI: Velocity Time IntegralXMean V: Mean

Strona 45 - Related Guidance Documents

5-15Chapter 5 Measurements And Calculations4. When the measurement is nished, its result is shown on the screen. XCirc.: Circumference of ObjectX

Strona 46 - -(0.069 lf)

14User ManualChapter 3 Starting DiagnosisPower Supply ...

Strona 47

5-16User ManualThe equations used for Trace measurement are as follows:, (N = 1,2… last point) , (N = 1,2… last point)Volume Measurement 3 DistanceThi

Strona 48 - Systematic Uncertainties

5-17Chapter 5 Measurements And Calculations2. Specify two points and measure the straight distance between them. The method for measuring a line is

Strona 49 - Training

5-18User Manual Calculations by Application Press the Calc button on the control panel.Things to note Before Taking Measurements Register PatientMak

Strona 50 - Environmental Protection

5-19Chapter 5 Measurements And Calculations Select Measurement ItemUse the Menu dial-button on the control panel. Return to Previous MenuAfter plac

Strona 51 - Introduction

5-20User ManualIn case of measuring distance, choose a measurement method from Line or Trace. When circumference and area are measured, choose a measu

Strona 52

5-21Chapter 5 Measurements And Calculations Limited TraceIf you specify a measurement range, a spectrum is traced automatically. It is enabled in th

Strona 53 - Specications

5-22User ManualTips!Things to consider for Doppler Spectrum Auto TraceThe state of a Doppler spectrum may affect measurement results. Please see the fo

Strona 54

5-23Chapter 5 Measurements And Calculations Vesl. Dist. (Vessel Distance)Measure the width of a blood vessel and calculate TAMV and Volume Flow. Ste

Strona 55 - Product Conguration

5-24User Manual Heart Rate Measurement HR (Heart Rate)You can calculate heart rates for a certain period of time.1. Select HR in the Measurement men

Strona 56

5-25Chapter 5 Measurements And Calculations Measurement MenuWhen the measurements for the selected items are complete, the measurements and gestation

Strona 57 - Control Panel Map

15Table of Contents Chapter 5 Measurements And CalculationsMeasurement Accuracy ...

Strona 58

5-26User ManualMeasurement Menu Item Mode Method UnitFetal BiometryGS All Distance cm, mmCRL All Distance cm, mmYS All Distance cm, mmBPD All Distance

Strona 59 - Chapter 2 Introduction

5-27Chapter 5 Measurements And CalculationsMeasurement Menu Item Mode Method UnitFetal OthersFoot All Distance cm, mmEar All Distance cm, mmMP All Di

Strona 60 - [Figure 2.4 keyboards]

5-28User Manual Automatic Calculation Some items in the measurement menu are automatically calculated based on measurements of other items.  HC This

Strona 61 - Adjusting Control Panel

5-29Chapter 5 Measurements And Calculations Calculating Estimated Fetal Weight (EFW)Estimated fetal weight is calculated automatically when the follo

Strona 62 - [Figure 2.5 Console]

5-30User ManualNOTE:XFor information on basic measurements, please refer to “Basic Measurements” and “Common Measurement Methods” in this chapter.XF

Strona 63 - [Figure 2.6 Rear Panel]

5-31Chapter 5 Measurements And CalculationsCardiac CalculationsNOTE:XCardiac measurement is an optional item for SONOACE R5. XSince the cardiac blo

Strona 64 - Power Connection Part

5-32User ManualMeasurement Menu Item Mode Method Unit2D MeasureAll All Continuouscm, mmIVSd All Distancecm, mmIVSs All Distancecm, mmLVDd All Distanc

Strona 65 - Probe Holder

5-33Chapter 5 Measurements And CalculationsMeasurement Menu Item Mode Method UnitMV (M)All Points M ContinuousD-E M Distancecm, mmE-F Slope M Veloc

Strona 66

5-34User ManualMeasurement Menu Item Mode Method UnitAoV SystolicAuto Trace PW Doppler Spectrum TraceLimited Trace PW Doppler Spectrum TraceManual Tra

Strona 67

5-35Chapter 5 Measurements And CalculationsMeasurement Menu Item Mode Method UnitTV RegurgAuto Trace PW Doppler Spectrum TraceLimited Trace PW Dopple

Strona 69 - Accessory

5-36User ManualMeasurement Menu Item Mode Method UnitTissue DopplerE-DT-A PW Time-Velocity-Timems & m/sPeak E’ PW Velocitym/sPeak A’ PW Velocitym/

Strona 70

5-37Chapter 5 Measurements And CalculationsNOTE:XFor information on basic measurements, please refer to “Basic Measurements” and “Common Measurement

Strona 71 - Starting Diagnosis

5-38User ManualMeasurement Menu Item Mode Method UnitRt./Lt. Mid CCA Same as aboveRt./Lt. Distal CCA Same as aboveRt./Lt. Bulb Same as aboveRt./Lt. Pr

Strona 72

5-39Chapter 5 Measurements And CalculationsUrology CalculationsThe measurement methods of each menu vary with the Volume Method set at Setup > Mea

Strona 73 - Power Supply

5-40User Manual 3 Distance Volume is calculated using three distance measurement results.Measurement Menu Item Mode Method UnitVolume FlowAuto Trace

Strona 74 - [Figure 3.1 Application]

5-41Chapter 5 Measurements And CalculationsMeasurement Menu Item Mode Method UnitResidual Vol.Pre All AllCalculated after distance measurementmlPre 1

Strona 75 - Probe Setting Change

5-42User ManualMeasurement Menu Item Mode Method UnitT-Z Prostate Vol. Same as aboveBladder Vol. Same as aboveRt./Lt. Kidney Vol.All AllCalculated aft

Strona 76 - Editing BodyMarker

5-43Chapter 5 Measurements And Calculations Sum of 20 DisksMeasure the prostate circumference and then use the trackball and the Set button to measu

Strona 77 - Save BodyMarkers

5-44User ManualFetal Echo CalculationsThe measurement method for each item is the same as for basic measurement. In addition, measurement items are si

Strona 78 - Calc Sequence

5-45Chapter 5 Measurements And CalculationsMeasurement Menu Item Mode Method Unit2D Echo Asc. Aorta All Distance cm, mmMPA All Distance cm, mmDuct.

Strona 79 - Patient Information

Chapter 1SafetyPurpose of Use ... 1-3Contraindications ...

Strona 80 - Registering a New Patient

5-46User ManualMeasurement Menu Item Mode Method UnitIVCSame as aboveDuct. VenSame as aboveAsc. AortaSame as aboveDesc. AortaSame as above Same as abo

Strona 81

5-47Chapter 5 Measurements And Calculations ReportCheck the measurement and calculation results in the report of each application. XPress the Repor

Strona 82

5-48User ManualViewing ReportXThe measurement and calculation items and their values are displayed. XWhen the report screen has more than two pages,

Strona 83 - [Figure 3.5 Search]

5-49Chapter 5 Measurements And Calculations Open Line TransferTransfer the report to another computer or printer. Press the Open Line Transfer on the

Strona 84

5-50User Manual Viewing Graph XThe graph of the recently measured values is drawn according to the gestation period (week) and fetal size (mm).XThe

Strona 85 - Modifying Patient Information

SAMSUNG MEDISON DIAGNOSTIC ULTRASOUND SYSTEMSONOACE R5User ManualVolume 2

Strona 87 - Diagnosis Modes

SAMSUNG MEDISON DIAGNOSTIC ULTRASOUND SYSTEMVersion 1.02EnglishSONOACE R5User Manual

Strona 89 - Information

5Table of Contents Table of Contents – Volume 2Chapter 6 Image ManagementsCINE / LOOP ...

Strona 91 - Frequency

6User ManualSetting DICOM (Optional) ...7-23Setting

Strona 92 - Harmonic

7Table of Contents Chapter 9 ProbesProbes ...

Strona 94 - 2D Mode Menu

Chapter 6Image ManagementsCINE / LOOP ... 6-3Annotating Images ...

Strona 96 - [Figure 4.2 2D Post]

6-3Chapter 6 Image Managements CINE / LOOPImages are automatically saved in the memory during scanning. The saved images can be useful when a patien

Strona 97 - [Figure 4.3 2D Post Edit]

6-4User Manual How to Review ImagesUse the trackball on the control panel. You can search for saved images in chronological order by rotating the Trac

Strona 98 - 2D Mode Softmenu

6-5Chapter 6 Image ManagementsNOTE:XWhen saving a Cine image, the current time at the top right of the screen stops displaying. When the image is sa

Strona 99 - Entering and Exiting M Mode

6-6User Manual Annotating Images TextYou can enter comments or text over an image. This function can be useful when the diagnosis area is dierentiat

Strona 100 - M Mode Menu

6-7Chapter 6 Image Managements AutotextThis function allows you to enter text automatically by using an abbreviation. In this way, you can enter text

Strona 101 - M Mode Softmenu

1-3Chapter 1 Safety Purpose of UseThe SONOACE R5 Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and 

Strona 102 - Entering and Exiting C Mode

6-8User ManualBodyMarker You can place a BodyMarker over an image. This function can be useful when the diagnosis area is dierentiated or displayed.[

Strona 103 - C Mode Menu

6-9Chapter 6 Image Managements Entering BodyMarker 1. Press the BodyMarker key on the control panel, BodyMarkers will appear on the user information

Strona 104 - User Manual

6-10User ManualIndicatorYou can place an Indicator over an image. This function can be useful when the diagnosis area is dierentiated or displayed.

Strona 105 - C Mode Softmenu

6-11Chapter 6 Image Managements Deleting Indicator Press the Clear button on the control panel and all Indicators displayed on the screen will be del

Strona 106 - PD Mode Screen

6-12User Manual Saving and Transferring ImagesSaving ImagesWARNING: You must enter the patient ID because all images are saved according to patient I

Strona 107 - PD Mode Softmenu

6-13Chapter 6 Image Managements Transferring Images in SONOVIEWSelect an exam in the Exam View screen to transfer all images in the exam, or select a

Strona 108 - PW Spectral Doppler Mode

6-14User Manual Printing and Recording ImagesPrinting ImagesPress the Print button on the control panel to print out a scanned image using a printer.

Strona 109 - Sample Volume

6-15Chapter 6 Image Managements SONOVIEWSONOVIEW is an integrated image management program in SONOACE R5. It allows you to save, view and delete imag

Strona 110 - [Figure 4.8 HPRF]

6-16User ManualExam ViewNOTE: If you select an icon on the Exam View screen, the name of icon will be displayed. Starting Exam ViewPress Exam Search

Strona 111 - Chapter 4 Diagnosis Modes

6-17Chapter 6 Image Managements Selecting ExamUse the trackball and the Set button to select an examination. To cancel the exam selection, press the

Strona 113 - Combined Modes

1-4User Manual Safety SignsPlease read this chapter before using the Samsung Medison ultrasound system. It is relevant to the ultrasound system, the

Strona 114 - Changing Menu

6-18User Manual1. Select exam(s) and press Export to bring up the Export window. 2. Select MO drive, Flash Memory, or ODD as the destination drive f

Strona 115 - Dual Mode

6-19Chapter 6 Image ManagementsTips!When selecting 2D Cine, an image and a video are saved together. The saved AVI file (*.avi) can be played by a vi

Strona 116 - 3D Mode

6-20User ManualWhen you have nished using the drive, you must use the Eject function from Storage Manager to remove the MO disk or disconnect the dri

Strona 117 - Acquiring a 3D Image

6-21Chapter 6 Image Managements Sending Exam This function uses the DICOM network. The DICOM Send button will not work if the DICOM function is not

Strona 118 - Viewing a 3D Image

6-22User Manual Printing ExamThis function uses printer(s) on the DICOM network. DICOM Print will not work if the DICOM function is not congured.1.

Strona 119 - 3D Image Optimization

6-23Chapter 6 Image Managements Exam InformationPatient names, IDs, and exam dates are displayed in the left-hand center of the Exam Review screen.

Strona 120 - Position

6-24User Manual Selecting imaages and Full Screen Image View  Selecting ImagesUse the trackball on the control panel to move the cursor over an imag

Strona 121 - Cut Mode

6-25Chapter 6 Image ManagementsXImages and Report : prints both images and reportXCopies: sets the number of copies3. Press Ok and Print screen wi

Strona 122 - [Figure 4.16 Multi planar]

6-26User Manual E-mailThis function sends e-mails with image attachments.1. Select image(s) and press E-mail on the Exam Review screen.2. Enter the

Strona 123

6-27Chapter 6 Image Managements ExportThis function saves the images in BMP, JPEG, TIFF or DICOM le formats.1. Press Export on the Exam Review scre

Strona 124

1-5Chapter 1 Safety Symbols Description Symbols DescriptionPower on Microphone PortPower o Probe portPower ON for part of the productProtected agai

Strona 125 - Calculations

6-28User Manual4. Use the trackball to move the cursor to another location and press the Set button to set the end point. XPress the Change button b

Strona 126

6-29Chapter 6 Image Managements LayoutUse this function to congure the number of images to show on the Exam Review screen. On the Exam Review screen

Strona 128 - Human Error

Chapter 7UtilitiesSetting System ... 7-3General ...

Strona 129 - Doppler Mode

Chapter 7Information ...7-35Utility ...

Strona 130 - Cursor Position

7-3Chapter 7 Utilities Setting System This mode is used for system settings. It does not aect image output. The setup may be modied depending on s

Strona 131 - PW Spectral Doppler Mode

7-4User Manual InstitutionEnter the name of the hospital/institution. LanguageThis sets the language to be used. English, Deutsch, Français, Italian

Strona 132 - [Table 5.1 Basic Measurement]

7-5Chapter 7 Utilities Boot up caps lock statusThis menu sets the initial status of Caps Lock after system boot-up. Its default value is ‘O’. This

Strona 133 - 2D Distance (Line)

7-6User ManualDisplay To set the information about images and related data, select the Display tab in the System menu. Auto FreezeAfter the preset ti

Strona 134 - 2D Distance (Trace)

7-7Chapter 7 Utilities Simultaneous ModeThis menu determines whether or not simultaneous mode is enabled in PW Spectral Doppler Mode. X‘O’: Select

Strona 135 - 2D Hip Joint

1-6User ManualLabels Phrases containing the words ‘warning’ and/or ‘caution’ are displayed on the product's surface in order to protect it.

Strona 136 - Doppler Velocity

7-8User ManualMisc.Select the Misc. tab in the System menu.[Figure 7.5 System – Misc.] OB Measure Shortcut KeySelect On to start a desired obstetrics

Strona 137 - Doppler Trace

7-9Chapter 7 Utilities ThumbnailTo use thumbnail list, set this item as ‘On’. Thumbnails of saved images are showed up on the right side of the scre

Strona 138 - 2D Ellipse

7-10User Manual Setting Measurements Select Measure in the Setup menu to set up measurement functions.1. Press the Setup key on the keyboard. Setup

Strona 139 - 22D Trace

7-11Chapter 7 Utilities GeneralSelect General tab. Measure Item Display CountSelect the number of measurement items to be between 1 and 4. If you se

Strona 140 - 1 Distance

7-12User Manual Doppler Select the Doppler tab in the General tab. [Figure 7.7 Measure - Doppler] Y-Axis UnitThis sets the vertical (Y) axis unit in

Strona 141 - Distance + Ellipse

7-13Chapter 7 UtilitiesTips!How to change Doppler setting during measurementPress the space bar after pressing the Calc button to bring up the Settin

Strona 142 - Measurement Operations

7-14User Manual ReportSelect the Report tab in the General tab.Select items to mark the header on the report. The header will be shown only the item h

Strona 143 - Softmenu

7-15Chapter 7 UtilitiesOBSelect the OB tab in the Measure menu.[Figure 7.10 Measure – OB] Display SD ValueThis menu sets whether SD (Standard Deviat

Strona 144 - Common Measurement Methods

7-16User ManualXAuxiliary Unit: This sets the auxiliary unit of fetal weight measurement. Unit sets the available auxiliary unit. For instance, set t

Strona 145 - Manual Trace

7-17Chapter 7 Utilities GraphThe user can select graphs to mark on the OB report or print. Press Graph button and select items. After selecting/dese

Strona 146 - Volume Flow Measurement

1-7Chapter 1 Safety Electrical SafetyThis equipment is categorized as a Class I device with Type BF applied parts.CAUTION: As for US requirement,

Strona 147 - Stenosis Measurement

7-18User Manual Backup / RestoreXBackupIf you want to back up a new user preset to MO Media, Flash Memory, HDD, or CD, press Backup/Restore on the s

Strona 148 - Heart Rate Measurement

7-19Chapter 7 UtilitiesFetal Echo Select the Fetal Echo tab in the Measure menu.[Figure 7.14 Measurement Setting– Fetal Echo]During obstetric measure

Strona 149 - Measurement Menu

7-20User ManualCardiac Select the Cardiac tab in the Measure menu.  2D Circumference and Area MethodSelect either the ‘Ellipse’ or ‘Trace’ caliper to

Strona 150

7-21Chapter 7 UtilitiesUrology Select the Urology tab in the Measure menu. [Figure 7.16 Measure– Urology] Volume MethodX3 Distance: The volume valu

Strona 151

7-22User Manual Predicted PSA Correction Factor The Predicted PSA Correction Factor can be changed. The default value is 0.12.Vascular Select the Vas

Strona 152 - AFI (Amniotic Fluid Index)

7-23Chapter 7 Utilities Setting DICOM (Optional)Select the DICOM tab in the Setup menu. This function is used to set up the DICOM server and other D

Strona 153 - GYN Calculations

7-24User ManualSetting DICOM InformationEnter the DICOM information for the product in use. For automatic DICOM transmission, select ‘Acquisition in P

Strona 154

7-25Chapter 7 Utilities AliasEnter the name of the DICOM server being used. AE TitleEnter the AE title of the DICOM server being used. Consult your

Strona 155 - Cardiac Calculations

7-26User Manual[Figure 7.19 DICOM Configuration-Store] Print Server InformationSelect PRINT under Services. Congure the Print Service using DICOM. NO

Strona 156

7-27Chapter 7 Utilities Empty DensitySpecify the background color of an image to print. Select Black or White. Min DensitySpecify the minimum brigh

Strona 157

1-8User ManualCAUTION:XThe system has been designed for 100-120VAC and 200-240VAC; you should select the input voltage of printer and VCR. Prior to c

Strona 158

7-28User Manual[Figure 7.20 DICOM Configuration - Print] Worklist Server SettingsSelect WORKLIST under Services. Congure the Modality Worklist Servic

Strona 159

7-29Chapter 7 Utilities Scheduled Station AE TitleSets the range of AE Title to fetch from the hospital’s Worklist server.XAll: Obtains the list of

Strona 160 - Vascular Calculations

7-30User ManualEditing the DICOM Server Information Click Edit in the setting DICOM window to edit the server information already entered.Deleting DIC

Strona 161

7-31Chapter 7 Utilities DICOM LogView, copy, or delete the DICOM log les.XView Selected File: Select the log le from ‘Select log les to copy’ an

Strona 162

7-32User Manual Setting OptionThe S/W serial No information of the system is shown in this window. You can select/cancel S/W options. A user cannot m

Strona 163 - Urology Calculations

7-33Chapter 7 Utilities Setting Peripheral DevicesThe following describes how to set up the video output type, video input type, printer, foot switc

Strona 164 - 3 Distance

7-34User ManualXHP Color LaserJet 3600XHP DeskJet 5940 XHP DeskJet 6540 XHP DeskJet 6940XHP OceJet J5780XHP LaserJet 1320XHP LaserJetP2015XHP

Strona 165 - Ellipsoid

7-35Chapter 7 Utilities InformationThe information menu displays information about the system S/W version. Select the Information tab in the Setup me

Strona 166

7-36User Manual Utility Press Utility key on the keyboard. The Utility menu appears at the left side of the screen. You can select a menu with the Men

Strona 167 - Sum of 20 Disks

7-37Chapter 7 Utilities Editing Biopsy guide lineNOTE:XNote that the Biopsy Guide Line cannot be edited when the Trapezoidal function is in use for

Strona 168 - Fetal Echo Calculations

1-9Chapter 1 Safety ESDElectrostatic discharge (ESD), commonly referred to as a static shock, is a naturally occurring phenomenon. ESD is most preva

Strona 169

7-38User Manual Starting and Finishing BiopsyNOTE: The biopsy guideline must be adjusted before a biopsy can be performed.1. Select On in the Biopsy

Strona 170

7-39Chapter 7 UtilitiesMonitor CalibrationSelect Monitor Calibration in Utility menu. Set the color or brightness of the image. [Figure 7.28 Monitor

Strona 171 - Report

7-40User ManualPresetPreset allows the user to adjust Image Settings and select a Setting to apply to the system.Select Preset in the utility menu.[Fi

Strona 172 - Exporting Report

7-41Chapter 7 UtilitiesMiscellaneousSelect Miscellaneous in the utility menu.[Figure 7.30 Miscellaneous] HistogramShows the brightness level of the s

Strona 173 - Open Line Transfer

7-42User Manual[Figure 7.31 Histogram]

Strona 174 - Viewing History

Chapter 8MaintenanceOperating Environment ...8-3System Maintenance ... 8-4Cleaning

Strona 176

8-3Chapter 8 Maintenance Operating EnvironmentWhen installing the console, please consider the following:CAUTION: Too large a difference in the gain

Strona 177

8-4User Manual System MaintenanceNOTE: The expected service life of this equipment is 7 years.Cleaning and disinfections Using inappropriate detergen

Strona 178

8-5Chapter 8 Maintenance DisinfectionsCAUTION: Use only recommended disinfectants on system surfaces.A disinfectant qualied by the FDA 510(k) proces

Strona 179 - Table of Contents – Volume 2

1-10User ManualEMC The testing for EMC(ElectroMagnetic Compatibility) of this system has been performed according to the international standard for EM

Strona 180 - Chapter 8 Maintenance

8-6User ManualFuse ReplacementThe power protection fuse protects the product from excess current. If the power monitoring protection circuit detects e

Strona 181 - Chapter 9 Probes

8-7Chapter 8 MaintenanceCleaning The Air FiltersThe air lters minimize the indraft of dust. Clean the air lter to ensure that a clogged lter does

Strona 182

8-8User ManualAccuracy CheckNOTE: The user must ensure that safety inspections are performed every 2 years according to the requirements of safety sta

Strona 183 - Image Managements

8-9Chapter 8 Maintenance Data MaintenanceCAUTION: You may lose information files on user settings or patients, because of shock on the product or int

Strona 184

8-10User ManualSoftwareChanges can be made to the product software to make it suitable for a particular user’s environment. However, these software ch

Strona 185 - [Figure 6.1 Reviewing Images]

Chapter 9ProbesProbes ... 9-3Ultrasound transmission Gel ...

Strona 187 - Speed (%)

Chapter 9 Probes9-3 ProbesThe probe is a device that sends and receives ultrasound for acquiring image data. It is also called a Transducer or Scanhe

Strona 188 - [Figure 6.2 Text Mode]

9-4User ManualProbe Application Biopsy HarmonicPulse InversionQ ScanCN2-8OBO O X OGynecologyAbdomenFetal HeartRenalCN4-9NeonatalX X X OPediatricVascul

Strona 189 - Exiting Text Mode

Chapter 9 Probes9-5 TI TableTI (Thermal Index) is displayed on the title of the screen and shows the estimate of the temperature increase in specic

Strona 190 - Starting BodyMarker Mode

1-11Chapter 1 Safety Approved Cables, Probes and Peripherals for EMC CablesCables connected to this product may aect its emissions; Use only the

Strona 191 - Exiting BodyMarker Mode

9-6User ManualUltrasound transmission GelUsing an inappropriate ultrasound gel may damage the probe. For proper transmission of the acoustic beam, onl

Strona 192 - Entering Indicator

Chapter 9 Probes9-7 Installing the Sheath1. Use sterile gloves.2. Unpack the sheath and ll it with acoustic coupling gel.3. Insert the probe into

Strona 193 - Exiting Indicator Mode

9-8User Manual Use and Infection Control of the ProbeWARNING: No neurosurgical treatments or examinations should be carried out on a patient with Creu

Strona 194 - Saving an Image with Text

Chapter 9 Probes9-9Cleaning and Disinfecting the ProbeUsing an inappropriate detergent or disinfectant may damage the probe.WARNING: XAlways use pro

Strona 195 - Chapter 6 Image Managements

9-10User ManualNamesDisinfectantsT-Spray IIT-SpraySani-Cloth HBSani-Cloth PlusSani-Cloth ActiveSeptiwipesClieanisept WipesSter-Bac BluTranseptic Spray

Strona 196 - Recording Images

Chapter 9 Probes9-11NamesDisinfectantsCidex OPA 2), 3), 6), 7)Cidex Plus 2), 7)Metricide 2), 7)Omnicide (28)Omnicide 14NSOmnicide - FG2NucleanWavicid

Strona 197 - [Figure 6.6 SONOVIEW]

9-12User ManualNamesDisinfectants CleanerVirkonSporoxSporox II GigaseptGigasept AF 3)Gigasept FFHibitanePeraSaveEnzolAlkazymeCidezymeKlenzymeIsopropyl

Strona 198 - [Figure 6.7 Exam View]

Chapter 9 Probes9-13NamesCleaner GelEthanol 75%MetrizymeMcKessonNatural ImageAquasonics 100 3)GE Ultrasound Contact GelClear ImageKendallScanWaveleng

Strona 199 - Exam Export

9-14User ManualTips!Symbols (1) Compatible but no EPA Registration (2) FDA 510(k) cleared (3) Has CE mark (4) Discontinued (5) Under Developm

Strona 200 - [Figure 6.8 Exam Export]

Chapter 9 Probes9-15Following is information about manufacturer (or Distributor) of Detergent, Disinfectant, and Ultrasound Gel. Product Manufacture

Strona 201 - Exam Restore

1-12User ManualImmunity test IEC 60601 Test level Compliance levelElectromagnetic environment -guidanceElectrostatic discharge (ESD)IEC 61000-4-2±6KV

Strona 202 - Using ODD

9-16User Manual CleaningCleaning is an important procedure that is carried out before disinfecting the probe. The probe must be cleaned after each use

Strona 203 - [Figure 6.9 DICOM Send]

Chapter 9 Probes9-17 DisinfectionOnly disinfect vaginal and rectal probes. A 10-6 reduction in pathogens should be reached following the disinfection

Strona 204 - [Figure 6.10 Exam Review]

9-18User Manual BiopsyA biopsy is an examination method that surgically extracts tissue from the patient for examination. The probe and the biopsy kit

Strona 205 - Exam Search

Chapter 9 Probes9-19Using the Biopsy KitWARNING: XVerify the condition of the biopsy needle before use. Do not use a bent biopsy needle.XThe biopsy

Strona 206 - Printing

9-20User Manual5. Insert the needle into the needle guide. Perform the puncture by sliding the needle through the groove in the guide until the needl

Strona 207 - DICOM Printing (Optional)

Chapter 9 Probes9-21Cleaning and Disinfecting Biopsy KitWash and disinfect the biopsy kit to reduce pathogens to the level of 10-6. Some components o

Strona 208 - Magnication

9-22User Manual Cleaning and Disinfection of plastic biopsy kit Cleaning 1. After use, remove the biopsy kit from the probe.2. Disassemble the biop

Strona 209 - Distance

Chapter 9 Probes9-23Assembling the Biopsy Kit C2-5Biopsy Guide (Plastic)  SpecicationModel nameBP-KIT-045ComponentBiopsy Bracket 1 EANeedle Guide(

Strona 210 - Ellipse

9-24User Manual3. Secure the Needle Guide Clip.4. Insert the needle into the needle guide for use.

Strona 211 - Closing SONOVIEW

Chapter 9 Probes9-25 C2-5Biopsy Guide (Metal)  SpecicationModel nameBP-KIT-042ComponentBiopsy Bracket 1 EANeedle Guide(16G) 6 EANeedle Guide(18G)

Strona 212

1-13Chapter 1 Safety Immunity testIEC 60601Test levelCompliance levelElectromagneticenvironment - guidanceConducted RFIEC 61000-4-63 Vrms150 kHzto 8

Strona 213 - Utilities

9-26User Manual3. Secure the Needle Guide Clip.4. Insert the needle into the needle guide for use.

Strona 214 - Chapter 7

Chapter 9 Probes9-27 C2-8Biopsy Guide  SpecicationModel nameBP-KIT-035ComponentBiopsy Bracket Needle Guide 1 EA 24 EAMaterialAcetal CopolymerAvail

Strona 215 - [Figure 7.1 System – General]

9-28User Manual3. Secure the Needle Guide Clip.4. Insert the needle into the needle guide for use.

Strona 216 - Language

Chapter 9 Probes9-29 CN2-8 Biopsy Guide  SpecicationModel nameBP-KIT-038ComponentAdapter 1EANeedle Guide(16G) 8 EANeedle Guide(18G) 8 EANeedle Gu

Strona 217 - Autotext Set

9-30User Manual3. Secure the Needle Guide Clip. 4. Insert the needle into the needle guide for use.

Strona 218 - Display

Chapter 9 Probes9-31 L5-12/60 Biopsy Guide  SpecicationModel nameBP-KIT-039ComponentAdapter 1 EANeedle Guide(16G) 8 EANeedle Guide(18G) 8 EANeed

Strona 219 - Simultaneous Mode

9-32User Manual3. Secure the Needle Guide Clip. 4. Insert the needle into the needle guide for use.

Strona 220 - [Figure 7.5 System – Misc.]

Chapter 9 Probes9-33 LE5-12 Biopsy Guide  SpecicationModel nameBP-KIT-040ComponentBiopsy Adapter 1 EAMaterialAcetal CopolymerAvailable Gauges16G,

Strona 221 - Trackball Speed

9-34User Manual3. Secure the Needle Guide Clip. 4. Insert the needle into the needle guide for use.

Strona 222 - Setting Measurements

Chapter 9 Probes9-35 LN5-12/40 Biopsy Guide  SpecicationModel nameBP-KIT-036ComponentAdapter 1 EANeedle Guide 24 EAMaterialAcetal CopolymerAvaila

Strona 224 - Doppler

1-14User Manual Recommended separation distances between portable and mobile RF communications equipment and the SONOACE R5This product is intended f

Strona 225 - Chapter 7 Utilities

9-36User Manual3. Secure the Needle Guide Clip.4. Insert the needle into the needle guide for use.

Strona 226 - [Figure 7.9 Measure - Report]

Chapter 9 Probes9-37 LN5-12 Biopsy Guide  Specication Model nameBP-KIT-040ComponentBiopsy Adapter 1 EA 16Ga disposable needle guide/probe cover K

Strona 227 - [Figure 7.10 Measure – OB]

9-38User Manual3. Secure the Needle Guide Clip. 4. Insert the needle into the needle guide for use.

Strona 228 - [Figure 7.11 New Preset]

Chapter 9 Probes9-39 EC4-9, EVN4-9 Biopsy guide SpecicationsModel nameBP-KIT-024ComponentBiopsy Adapter 1 EASheath 2 EA MaterialStainlessAvailab

Strona 230 - [Figure 7.13 Backup/Restore]

SAMSUNG MEDISON DIAGNOSTIC ULTRASOUND SYSTEMSONOACE R5Reference Manual

Strona 232 - Cardiac

SAMSUNG MEDISON DIAGNOSTIC ULTRASOUND SYSTEMVersion 1.02EnglishSONOACE R5Reference Manual

Strona 234 - Vascular

PART 1. OB REFERENCEOB REFERENCE 6Gestational aGe Formula and table list 6ABDOMINAL CIRCUMFERENCE (AC)KOREAN 6HADLOCK 6HANSMANN 7MERZ 8JEANTY 11SHIN

Strona 235 - [Figure 7.18 Setup - DICOM]

1-15Chapter 1 Safety Avoiding Electromagnetic InterferenceTypical interference on Ultrasound Imaging Systems varies depending on Electromagnetic p

Strona 236 - Store Server Information

Reference Manual 2 THORAX TRANSVERSE DIAMETER (TTD)HANSMANN 52GESTATIONAL SAC (GS)KOREAN 53HANSMANN 53HELLMAN 54NYBERG 54TOKYO 54OCCIPITAL FRONTAL D

Strona 237

Reference Manual 3 UMBILICAL ARTERY (UMA)-RESISTANCE INDEX (RI)SHINOZUKA 84UMBILICAL ARTERY (UMA)- PULSATILITY INDEX (PI)SHINOZUKA 85ANTERIOR POSTER

Strona 238 - Print Server Information

Reference Manual 4 cardioloGy m mode 98Left Ventricle 98LV Vol. d (LV Volume Diastolic) 98LV Vol. s (LV Volume Systolic) 99Stroke Volume (SV) 9

Strona 239

Reference Manual 5 PART 3. ACOUSTIC POWER TABLESACOUSTIC POWER TABLES 109iec tables 109Symbols and Denitions 109Explanatory Notes 110C2-4/20 11

Strona 240 - Worklist Server Settings

Reference Manual 6 OB ReferenceGestational Age Formula and Table List Abdominal Circumference (AC) : KOREANGA TableY.G Park.“ The Standardization

Strona 241 - Start Date

Reference Manual 7 Fetal Growth TableHadlock, F., Deter, R.L., Harrist, R.B., Park, S.K. “Estimating Fetal Age: Computer-Assisted Analysis of Multip

Strona 242 - DICOM Log

Reference Manual 8 Fetal Growth TableHansmann, Hackeloer, Staudach, Wittman “Ultrasound Diagnosis in Obstetrics and Gynecology” Springer- Verlag, Ne

Strona 243

Reference Manual 9 Age (w) GA (wd) Min (wd) Max (wd) Age (w) GA (wd) Min (wd) Max (wd)9.40 15w5d 14w3d 17w1d 24.20 29w6d 27w6d 31w5d9.60 15w6d 14w4d

Strona 244 - [Figure 7.22 Setup - Option]

Reference Manual 10 Age (w) GA (wd) Min (wd) Max (wd) Age (w) GA (wd) Min (wd) Max (wd)18.60 24w4d 22w6d 26w1d 33.40 38w4d 36w2d 40w6d18.80 24w5d 23

Strona 245 - Setting Peripheral Devices

Reference Manual 11 Abdominal Circumference (AC) : JEANTYFetal Growth TableHansmann, Hackeloer, Staudach, Wittman. “Ultrasound Diagnosis in Obstet

Strona 246 - Printer 2

1-16User Manual Is the interference intermittent or constant? Does the interference show up only with one transducers operating at the same frequenc

Strona 247 - [Figure 7.24 Information]

Reference Manual 12 AC (cm) GA (wd) ±SD (wd) AC (cm) GA (wd) ±SD (wd)19.5 24w6d 1w4d 31.5 39w0d 2w1d20.0 25w3d 1w4d 32.0 39w6d 2w1d20.5 25w6d 1w4d 3

Strona 248 - [Figure 7.26 Biopsy menu]

Reference Manual 13 Age (W)Growth (cm)Min (cm)Max (cm)Age (W)Growth (cm)Min (cm)Max (cm)24 18.93 17.48 20.39 40 34.04 31.35 36.7325 19.98 18.4

Strona 249 - [Figure 7.27 Biopsy Edit]

Reference Manual 14 AC (cm) GA (wd) ±SD (wd) AC (cm) GA (wd) ±SD (wd)18.80 23w0d 02w3d 33.10 37w0d 04w2d19.90 24w0d 02w4d 33.80 38w0d 04w2d21.00 25w

Strona 250

Reference Manual 15 Abdominal Circumference (AC) : CFEFFetal Growth TableJ.Créquat, M. Duyme, G. BrodatyBiométrie 2000. Tables de croissance foeta

Strona 251 - Monitor Calibration

Reference Manual 16 Biparietal Diameter (BPD) : HANSMANNGA TableHansmann, Hackeloer, Stauch, Wittman “Ultrasound Diagnosis in Obsterics and Gyneco

Strona 252 - [Figure 7.29 Preset]

Reference Manual 17 Fetal Growth TableHansmann, Hackeloer, Stauch, Wittman “Ultrasound Diagnosis in Obsterics and Gynecology” Springer-Verlag, New Y

Strona 253 - Histogram

Reference Manual 18 Min Range : 12 wMax Range : 40 wStandard Deviation : 2SD = 0.6 cmMin Range (cm) Max Range (cm) SD (cm)0 100 0.6 Biparietal Diam

Strona 254 - [Figure 7.31 Histogram]

Reference Manual 19 BPD (cm) GA (wd) Min (wd) Max (wd) BPD (cm) GA (wd) Min (wd) Max (wd)5.40 21w4d 19w5d 23w4d 9.50 36w6d 34w4d 39w2d5.50 21w6d 20w

Strona 255 - Maintenance

Reference Manual 20 BPD (cm) GA (wd) Min (wd) Max (wd) BPD (cm) GA (wd) Min (wd) Max (wd)1.70 11w1d 08w3d 13w5d 6.00 23w4d 20w6d 26w1d1.80 11w2d 08w

Strona 256

Reference Manual 21 Biparietal Diameter (BPD) : SABBAGHAGA TableSabbagha, R.E., et. al. “Standardization of Sonar Cephalometry and Gestational Age

Strona 257 - Operating Environment

1-17Chapter 1 Safety Mechanical SafetyMoving the EquipmentWARNING: Be extra careful when transporting it. Careless transportation of the product m

Strona 258 - Cleaning

Reference Manual 22 BPD (cm) GA (wd) Min (wd) Max (wd) BPD (cm) GA (wd) Min (wd) Max (wd)3.00 15w1d 14w3d 15w6d 6.90 27w4d 26w1d 29w0d3.10 15w3d 14w

Strona 259 - Disinfections

Reference Manual 23 Biparietal Diameter (BPD) : OSAKAGA TableOsaka University Method 1989, 3 by Univ. Of OsakaBPD (cm)G.A (wd)BPD (cm)G.A (wd)BPD

Strona 260 - [Figure 8.1 Fuse replacement]

Reference Manual 24 GA (wd) Growth (cm) ±SD (cm) GA (wd) Growth (cm) ±SD (cm)11w6d 2.04 0.21 27w0d 6.95 0.3412w0d 2.09 0.21 27w1d 6.99 0.3412w1d 2.1

Strona 261 - Cleaning The Air Filters

Reference Manual 25 GA (wd) Growth (cm) ±SD (cm) GA (wd) Growth (cm) ±SD (cm)17w6d 4.15 0.27 33w0d 8.43 0.3718w0d 4.20 0.27 33w1d 8.46 0.3718w1d 4.2

Strona 262 - Accuracy Check

Reference Manual 26 GA (wd) Growth (cm) ±SD (cm) GA (wd) Growth (cm) ±SD (cm)23w6d 6.05 0.32 39w0d 9.30 0.3924w0d 6.09 0.32 39w1d 9.31 0.3924w1d 6.1

Strona 263 - Data Maintenance

Reference Manual 27 BPD (cm) G.A (wd) Min (wd) Max (wd) BPD (cm) G.A (wd) Min (wd) Max (wd)4.70 20w2d 18w5d 22w0d 8.20 32w5d 29w4d 36w1d4.80 20w4d 1

Strona 264 - Software

Reference Manual 28 BPD (cm) G.A (wd) Min (wd) Max (wd) BPD (cm) G.A (wd) Min (wd) Max (wd)2.70 13w6d 12w6d 15w0d 6.20 24w2d 22w1d 26w5d2.80 14w1d 1

Strona 265 - Chapter 9

Reference Manual 29 Age (W)Growth (cm)Min (cm)Max (cm)Age (W)Growth (cm)Min (cm)Max (cm)18 4.19 3.86 4.52 34 8.76 8.31 9.2119 4.54 4.20 4.88 35

Strona 266

Reference Manual 30 Biparietal Diameter (BPD) : ASUM (SCW)Fetal Growth TableAustralasian Society for Ultrasound in MedicinePolicies and Statements

Strona 267 - Probe List

Reference Manual 31 Age (W)Growth (cm)Min (cm)Max (cm)Age (W)Growth (cm)Min (cm)Max (cm)14 2.71 2.44 2.99 30 7.58 7.12 8.0415 3.09 2.79 3.38

Strona 268

1-18User ManualSafety NoteCAUTION:XDo not press the control panel excessively.XNever attempt to modify the product in any way. XCheck the operatio

Strona 269 - TI Table

Reference Manual 32 Crown-Rump Length (CRL) : HANSMANNGA TableHansmann, Hackeloer, Staudach, Wittman. “Ultrasound Diagnosis in Obsterics and Gynec

Strona 270 - Ultrasound transmission Gel

Reference Manual 33 Fetal Growth TableHansmann, Hackeloer, Staudach, Wittmann. “Ultrasound Diagnosis in Obstetrics and Gynecology.” Springer-Verlag,

Strona 271 - Installing the Sheath

Reference Manual 34 Fetal Growth TableN.Shinozuka. “Fetal biometry standard values” http://www.shinozuka.com, 1996Age (wd)Growth (cm)Min (cm)Max

Strona 272 - Electric Shocks

Reference Manual 35 CRL (cm) GA (wd)-2SD (wd)+2SD (wd)CRL (cm) GA (wd)-2SD (wd)+2SD (wd)0.2 5w5d 5w2d 6w1d 6.2 12w4d 11w4d 13w4d0.3 5w6d 5w3d 6w3d 6

Strona 273

Reference Manual 36 CRL (cm) GA (wd)-2SD (wd)+2SD (wd)CRL (cm) GA (wd)-2SD (wd)+2SD (wd)4.4 11w1d 10w2d 12w1d 10.4 16w2d 14w7d 17w4d4.5 11w2d 10w3d

Strona 274

Reference Manual 37 Fetal Growth TableOsaka University Method 1989, 3 by Univ. Of OsakaAge (w) Growth (cm) Min (cm) Age (w) Growth (cm) Min (cm)6 0.

Strona 275 - 2), 3), 6), 7)

Reference Manual 38 CRL (cm) GA (wd) CRL (cm) GA (wd) CRL (cm) GA (wd) CRL (cm) GA (wd)0.80 6w6d 2.30 8w5d 3.80 10w3d 5.30 12w0d0.90 7w0d 2.40 8w6d

Strona 276

Reference Manual 39 Max Range : 8.20 cmStandard Deviation :Min Range (w) Max Range (w) ±2SD (w)12 18 1.3818 24 1.8024 30 2.0830 36 2.9636 42 3.12Fet

Strona 277

Reference Manual 40 FL (cm)GA (wd)±SD (wd)FL (cm)GA (wd)±SD (wd)FL (cm)GA (wd)±SD (wd)2.60 17w6d 01w3d 5.00 26w6d 01w5d 7.40 37w2d 02w1d2.70 18w2d 0

Strona 278

Reference Manual 41 FL (cm) GA (wd) ±SD (wd) FL (cm) GA (wd) ±SD (wd)2.80 19w0d 1w3d 6.30 33w0d 2w6d3.10 20w0d 1w4d 6.50 34w0d 3w0d3.40 21w0d 1w4d 6

Strona 279

1-19Chapter 1 Safety Safety Note for MonitorWhen adjusting the height or position of the monitor, be careful of the space in the middle of the mon

Strona 280

Reference Manual 42 Femur Length (FL) : JEANTYGA TableJeanty, P., Rodesch, F., Delbeke, D., Dumont, J. “Estimation of Gestational Age from Measure

Strona 281 - [Figure 9.1 Disinfection]

Reference Manual 43 FL (cm) GA (wd) Min (wd) Max (wd) FL (cm) GA (wd) Min (wd) Max (wd)2.80 18w6d 17w6d 19w6d 5.40 30w2d 28w5d 31w6d2.90 19w1d 18w1d

Strona 282 - Biopsy KIT Components

Reference Manual 44 Femur Length (FL) : OSAKAGA TableOsaka University Method 1989, 3 by Univ. Of OsakaFL (cm) GA (wd) FL (cm) GA (wd) FL (cm) GA (

Strona 283 - Biopsy Procedure

Reference Manual 45 Age (wd) Growth (cm) ±SD (cm) Age (wd) Growth (cm) ±SD (cm)15w5d 1.79 0.22 29w2d 5.28 0.2815w6d 1.83 0.22 29w3d 5.31 0.2816w0d 1

Strona 284 - Needle Guide Alignment

Reference Manual 46 Age (wd) Growth (cm) ±SD (cm) Age (wd) Growth (cm) ±SD (cm)21w5d 3.49 0.25 35w2d 6.41 0.3021w6d 3.53 0.25 35w3d 6.43 0.3022w0d 3

Strona 285 - Disinfection

Reference Manual 47 Femur Length (FL) : CHITTYGA TableAltman DG, Chitty LS: New Charts for ultrasound dating of pregnancy. Ultrasound in Obstetric

Strona 286

Reference Manual 48 Age (W)Growth (cm)Min (cm) Max (cm) Age (W)Growth (cm)Min (cm) Max (cm)16 2.03 1.77 2.28 32 6.12 5.74 6.4917 2.33 2.07 2.59 33 6

Strona 287 - C2-5Biopsy Guide (Plastic)

Reference Manual 49 Age (wd) Growth (cm) Age (wd) Growth (cm) Age (wd) Growth (cm)18w0d 2.69 29w0d 5.63 40w0d 7.4719w0d 2.98 30w0d 5.80 41w0d 7.4820

Strona 288

Reference Manual 50 Femur Length (FL) : CFEFFetal Growth TableJ.Créquat, M. Duyme, G. BrodatyBiométrie 2000. Tables de croissance foetale par le C

Strona 289 - C2-5Biopsy Guide (Metal)

Reference Manual 51 Age (W)Growth (cm)Min (cm) Max (cm) Age (W)Growth (cm)Min (cm) Max (cm)17 3.11 2.61 3.61 32 7.62 6.52 8.7218 3.46 3.02 3.90 33 7

Strona 290

1-20User Manual Biological SafetyFor more safety information on probes and biopsy, please refer to Chapter 9 ‘Probes.’WARNING:XUltrasound waves may

Strona 291 - C2-8Biopsy Guide

Reference Manual 52 Fetal Growth TableNorio Shinozuka, Haruo Masuda, Hideyuki Kagawa, and Yuji Taketani. Department of Obstetrics and Gynecoogy, Fac

Strona 292

Reference Manual 53 Fetal Growth TableHansmann, Hackeloer, Staudach, Wittman “Ultrasound Diagnosis in Obstetrics and Gynecology” Springer-Verlag, Ne

Strona 293 - CN2-8 Biopsy Guide

Reference Manual 54 Gestational Sac (GS) : HELLMANGA TableHellman LM, Kobayashi M, Fillisti l, et al. Growth and development of the human fetus pr

Strona 294

Reference Manual 55 Occipital Frontal Diameter (OFD) : HANSMANNGA TableOFD (cm) G.A (wd) ±SD (wd) OFD (cm) G.A (wd) ±SD (wd)3.10 14w0d 01w3d 9.10

Strona 295 - L5-12/60 Biopsy Guide

Reference Manual 56 Occipital Frontal Diameter (OFD) : ASUM (SCW)Fetal Growth TableAustralasian Society for Ultrasound in MedicinePolicies and Sta

Strona 296

Reference Manual 57 Occipital Frontal Diameter (OFD) : MERZGA TableOFD (cm)GA (wd)OFD (cm)GA (wd)OFD (cm)GA (wd)OFD (cm)GA (wd)3.00 12w5d 5.60 19w

Strona 297 - LE5-12 Biopsy Guide

Reference Manual 58 Outer Ocular Distance (OOD) : JENATYGA TableJeanty, P., Rodesch, F., Delbeke, D., Dumont, J. “Estimation of Gestational Agefro

Strona 298

Reference Manual 59 Humerus (cm) GA (wd) ±SD (wd) Humerus (cm) GA (wd) ±SD (wd)1.30 13w4d 02w5d 4.30 25w5d 02w6d1.40 13w6d 02w5d 4.40 26w1d 02w6d1.5

Strona 299 - LN5-12/40 Biopsy Guide

Reference Manual 60 Age (W)Growth (cm)Min (cm) Max (cm) Age (W)Growth (cm)Min (cm) Max (cm)23 3.80 3.30 4.20 37 6.20 5.70 6.7024 4.00 3.50 4.50 38 6

Strona 300

Reference Manual 61 Fetal Growth TableMerz, Eberband, “Ultrasonic Mensuration of Fetal Limb Bones in the Second and Third Trimesters.” J Clin Ultras

Strona 301 - LN5-12 Biopsy Guide

1-21Chapter 1 Safety to the focal point, attenuation in the body, and ultrasound exposure time. Exposure time is an especially useful variable, beca

Strona 302

Reference Manual 62 Fetal Growth TableOsaka University Method 1989, 3 by Univ. Of OsakaAge (W) Growth (cm) ±SD (cm) Age (W) Growth (cm) ±SD (cm)13 1

Strona 303 - EC4-9, EVN4-9 Biopsy guide

Reference Manual 63 Tibia (TIB) : JENATYGA TableJeanty P, Rodesch F, Delbeke D, Mumont J, “Estimation of Gestational Age from Measurements of Feta

Strona 304

Reference Manual 64 Fetal Growth TableJeanty, P., Romero, R. “Obstetrical Ultrasound.” New York, McGraw-hill, 1983, p.192Age (W)Growth (cm)Min (cm)

Strona 305

Reference Manual 65 Age (W)Growth (cm)Min (cm) Max (cm) Age (W)Growth (cm)Min (cm) Max (cm)15 1.50 1.20 1.90 30 5.00 4.60 5.5016 1.80 1.40 2.10 31 5

Strona 306

Reference Manual 66 Cerebellum : CHITTYGA TableChitty LS, Altman DG, et al. : Charts of fetal size:2. Head Measurements Bri. J Obstet & Gyn. V

Strona 307

Reference Manual 67 Ulna : JEANTYGA TableJeanty P, Rodesch F, Delbeke D, Mumont J, “Estimation of Gestational Age from Measurements of Fetal Long

Strona 308

Reference Manual 68 Fetal Growth TableJeanty P, Rodesch F, Delbeke D, Mumont J, “Estimation of Gestational Age from Measurements of Fetal Long Bones

Strona 309

Reference Manual 69 Head Circumference (HC) : KOREANGA TableY.G Park.“ The Standardization of Fetal body parts according to the normal Korean Gest

Strona 310 - Reference Manual 2

Reference Manual 70 Fetal Growth TableHansmann, Hackeloer, Staudach, Wittman “Ultrasound Diagnosis in Obstetrics and Gynecology” Springer- Verlag, N

Strona 311 - CARDIOLOGY REFERENCE 95

Reference Manual 71 Fetal Growth TableHadlock, F., Deter, R.L., Harrist, R.B., Park, S.K. “Estimating Fetal Age: Computer-Assisted Analysis of Multi

Strona 312 - FETAL ECHO REFERENCE 107

1-22User ManualOutput has direct impact on acoustic intensity. Once the application has been established, the output control can be used to increase o

Strona 313 - ACOUSTIC POWER TABLES 109

Reference Manual 72 HC (cm) GA (wd) Min (wd) Max (wd) HC (cm) GA (wd) Min (wd) Max (wd)12.60 15w6d 14w3d 17w1d 27.40 28w4d 26w4d 30w3d12.80 16w0d 14

Strona 314 - OB Reference

Reference Manual 73 HC (cm) GA (wd) Min (wd) Max (wd) HC (cm) GA (wd) Min (wd) Max (wd)21.00 22w3d 20w5d 24w1d 35.80 39w4d 37w1d 41w6d21.20 22w4d 20

Strona 315 - Reference Manual 7

Reference Manual 74 HC (cm) GA (wd) Min (wd) Max (wd) HC (cm) GA (wd) Min (wd) Max (wd)12.50 16w2d 15w1d 17w3d 25.00 26w5d 24w6d 28w6d13.00 16w4d 15

Strona 316 - Reference Manual 8

Reference Manual 75 Head Circumference (HC) : CHITTY (M)GA TableChitty LS, Altman DG, et al.: Charts of fetal size: 2. Head measurements. Brit. J.

Strona 317 - Reference Manual 9

Reference Manual 76 Age (W)Growth (cm)Min (cm) Max (cm) Age (W)Growth (cm)Min (cm) Max (cm)19 16.43 15.40 17.46 35 32.33 30.83 33.8320 17.69 16.63 1

Strona 318 - Reference Manual 10

Reference Manual 77 Age (W) Growth (mm) ±SD (mm) Age (W) Growth (mm) ±SD (mm)17 141.00 15.00 33 300.00 25.0018 151.00 20.00 34 305.00 25.0019 160.00

Strona 319 - Reference Manual 11

Reference Manual 78 Fetal (cm2)GA (wd)Fetal (cm2)GA (wd)Fetal (cm2)GA (wd)Fetal (cm2)GA (wd)9.00 16w1d 30.00 24w0d 51.00 30w1d 72.00 35w5d10.00

Strona 320 - Reference Manual 12

Reference Manual 79 Age (wd)Growth (cm2)±SD (cm2) Age (wd)Growth (cm2)±SD (cm2) 16w5d 10.70 1.70 29w6d 50.20 5.8016w6d 11.00 1.80 30w0d 50.80 5.801

Strona 321 - Reference Manual 13

Reference Manual 80 Age (wd)Growth (cm2)±SD (cm2) Age (wd)Growth (cm2)±SD (cm2) 22w5d 25.90 3.30 35w6d 72.70 8.4022w6d 26.40 3.30 36w0d 73.20 8.402

Strona 322 - Reference Manual 14

Reference Manual 81 Fetal Growth TableYarkoni, S., Schmidt, W., Jeanty, P. et. al. (1985) Clavicle measurement: A new biometric parameter for fetal

Strona 323 - Reference Manual 15

1-23Chapter 1 Safety Receiver ControlsReceiver controls are used by the operator to improve image quality. These controls have no eect on output.

Strona 324 - Reference Manual 16

Reference Manual 82 Radius Length (RAD) : MERZFetal Growth Table[Fetal Limb Biometry] (Raiology 1083:147:602)Table Data : 95 percentile data form

Strona 325 - Reference Manual 17

Reference Manual 83 MAD (cm)G.A (wd)MAD (cm)G.A (wd)MAD (cm)G.A (wd)MAD (cm)G.A (wd)5.20 22w0d 7.00 27w2d 8.80 32w2d 10.60 38w5d5.30 22w2d 7.10 27w3

Strona 326 - Reference Manual 18

Reference Manual 84 Age (wd) 10% 90% Age (wd) 10% 90%28w3d 0.80 0.95 39w3d 0.68 0.8529w3d 0.80 0.94 40w3d 0.66 0.8430w3d 0.79 0.94 41w3d 0.64 0.8331

Strona 327 - Reference Manual 19

Reference Manual 85 Umbilical Artery (UmA)- Pulsatility Index (PI) : SHINOZUKAFetal Growth TableN.Shinozuka & H.Kagawa 1996. http://www.shinoz

Strona 328 - Reference Manual 20

Reference Manual 86 Anterior Posterior Abdominal Diameter (APD): BESSISGA TableThe data are those provided by Dr. Bessis to M. Le Bel. (Same as SI

Strona 329 - Reference Manual 21

Reference Manual 87 Head Circumference (HC) / Abdominal Circumference (AC): CampbellFetal Growth TableCampbell HC/AC Fetal Growth Tabel for Uranus

Strona 330 - Reference Manual 22

Reference Manual 88 Method using (BPD, FL, FTA)Osaka university [grams]FW (BPD,FL,FTA) = 1.256*BPD³+3.5065*FTA*FL+6.3 Method using (BPD, APTD, TT

Strona 331 - Reference Manual 23

Reference Manual 89 Method using (BPD, AC, FL)Hadlock 2 Method [grams][Equation]Log10 (EFW) = (0.0316 x BPD) + (0.0457 x AC) + (0.1623 x FL) - (0.

Strona 332 - Reference Manual 24

Reference Manual 90 Method using (BPD, APTD, TTD)Tokyo 1 university Method [grams][Equation]1.73 x BPD3 + 28 x APTD x TTD - 217[Reference]Jpn J Me

Strona 333 - Reference Manual 25

Reference Manual 91 Estimate Fetal Weight (EFW) : DOUBILETFetal Growth TableImproved Birth Weight Table for Neonates Developed from Gestations Dat

Strona 335 - Reference Manual 27

1-24User ManualSome factors might create articially high or low thermal index readings e.g. presence of uid or bone, or the ow of blood. A highly a

Strona 336 - Reference Manual 28

Reference Manual 92 Age (W) Growth (g) ±SD (g) Age (W) Growth (g) ±SD (g)18w4d 254 39 30w5d 1687 20018w5d 263 40 30w6d 1711 20218w6d 271 41 31w0d 17

Strona 337 - Reference Manual 29

Reference Manual 93 Age (W) Growth (g) ±SD (g) Age (W) Growth (g) ±SD (g)24w6d 822 108 37w0d 2767 31725w0d 839 110 37w1d 2791 32025w1d 857 112 37w2d

Strona 338 - Reference Manual 30

Reference Manual 94 Age (W)Growth (g)Min (g) Max (g) Age (W)Growth (g)Min (g) Max (g)22 478 398 559 38 3236 2686 378623 568 471 665 39 3435 2851 401

Strona 339 - Reference Manual 31

Reference Manual 95 Cardiology ReferenceCardiology 2D BSA (Body Surface Area)BSA can be calculated by entering patient’s weight and height in New

Strona 340 - Reference Manual 32

Reference Manual 96 Ejection Fraction (EF)where, EDV : End Diastolic Volume, ESV : End Systolic Volume Reference : Harvey Feigenbaum, “Echocardi

Strona 341 - Reference Manual 33

Reference Manual 97 LV Masswhere, A1 : short axis area, Epicardinal (cm2), A2 : short axis area, Endocardinal (cm2), L1 : long axis Epicardinal le

Strona 342 - Reference Manual 34

Reference Manual 98 Cardiology M mode Left VentricleReference : Harvey Feigenbaum, “ Echocardiography”, 1995 fth editionReference : Harvey Feigenb

Strona 343 - Reference Manual 35

Reference Manual 99 LV Vol. s (LV Volume Systolic)TeichholzReference : Teichholz, L.E., Kreulen, T., Herman, M.V., et. al. “Problems in echocardio

Strona 344 - Reference Manual 36

Reference Manual 100 MV (Mitral Valve)Definition for the Mitral ValveD : end of systolic, immediately before the opening of the Mitral ValveE :

Strona 345 - Reference Manual 37

Reference Manual 101 Cardiology C mode AoV Regurg, MV Regurg, TV Regurg, PV RegurgPISA (Proximal Isovelocity Surface Area)-Radius<Figure - Regur

Strona 346 - Reference Manual 38

1-25Chapter 1 Safety  Mechanical and Thermal indices Display Precision and AccuracyThe Mechanical and Thermal Indices on the system are precise to

Strona 347 - Reference Manual 39

Reference Manual 102 Eective Regurgitant Orice (ERO)Reference: Oh, J.K., Seward, J.B., Tajik, A.J. The Echo Manual. Boston:Little, Brown and Compa

Strona 348 - Reference Manual 40

Reference Manual 103 VTI (Velocity Time Integral)ACC (Acceleration), DEC (Deceleration)AT :Acceleration Time, DT : Deceleration Time, E point : Ear

Strona 349 - Reference Manual 41

Reference Manual 104 Diastolic VelocityVelocity measured at diastole.Systolic VelocityVelocity measured at systole.Artrial Reversal Velocity (A. Rev

Strona 350 - Reference Manual 42

Reference Manual 105 %STAReference : Jacob, Normaan M et, al., “ Duplex Carotid Sonography : Crieteria for Stenosis, Accuracy, and Pitfalls,” Radio

Strona 351 - Reference Manual 43

Reference Manual 106 Volume Flow (Area) Volume Flow (Dist.) Prostate Vloume (3 Distances) Prostate Vloume (3 Distances x Factor) Prostate Vlo

Strona 352 - Reference Manual 44

Reference Manual 107 Fetal Echo Reference Stroke Volume (SV)where, EDV : End Diastolic Volume, ESV : End Systolic Volume Cardiac Output (CO) E

Strona 353 - Reference Manual 45

Reference Manual 108 Cubed Reference : Pombo, J.F., et. al. “Left Ventricular Volumes and Ejection Fractioin by Echocardiography.” Circulation, Vol.

Strona 354 - Reference Manual 46

Reference Manual 109 Acoustic Power TablesIEC Tables Symbols and DenitionsStandard 60601-2-37 published by the International Electrotechnical Co

Strona 355 - Reference Manual 47

Reference Manual 110 Pr at max. Ipi the PEAK-RAREFACTIONAL ACOUSTIC PRESSURE at the point where the free -eld (non-attenuated), PULSE INTENSITY IN

Strona 356 - Reference Manual 48

Reference Manual 111 C2-4/20C2-4/20: 2D & 2D+M modeIndex Label M.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index Value0.862 0.8

Strona 357 - Reference Manual 49

1-26User Manualdierent tissues and organs have dissimilar attenuation characteristics. In water, there is almost no attenuation. In the body, and par

Strona 358 - Reference Manual 50

Reference Manual 112 C2-4/20: Pulsed Doppler & 2D+Pulsed DopplerIndex LabelM.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index Val

Strona 359 - Reference Manual 51

Reference Manual 113 C2-5: 2D+Color, Triple, CPA & Triple CPAIndex Label M.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index Value

Strona 360 - Reference Manual 52

Reference Manual 114 C2-8C2-8: 2D & 2D+M modeIndex Label M.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index Value 1.200 1.080 0.

Strona 361 - Reference Manual 53

Reference Manual 115 C2-8: Pulsed Doppler & 2D+Pulsed DopplerIndex Label M.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index Value

Strona 362 - Reference Manual 54

Reference Manual 116 CN2-8: 2D+Color, Triple, CPA & Triple CPAIndex Label M.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index Valu

Strona 363 - Reference Manual 55

Reference Manual 117 CN4-9CN4-9: 2D & 2D+M modeIndex Label M.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index Value 1.110 0.767

Strona 364 - Reference Manual 56

Reference Manual 118 CN4-9: Pulsed Doppler & 2D+Pulsed DopplerIndex LabelM.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index Value

Strona 365 - Reference Manual 57

Reference Manual 119 L5-12/60: 2D+Color, Triple, CPA & Triple CPAIndex Label M.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index V

Strona 366 - Reference Manual 58

Reference Manual 120 LE5-12LE5-12: 2D & 2D+M modeIndex Label M.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index Value 1.30 0.695

Strona 367 - Reference Manual 59

Reference Manual 121 LE5-12: Pulsed Doppler & 2D+Pulsed DopplerIndex Label M.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index Val

Strona 368 - Reference Manual 60

1-27Chapter 1 Safety  Focal no.More focal zones may change both the TI and MI by changing frame rate or focal depth automatically. Lower frame rate

Strona 369 - Reference Manual 61

Reference Manual 122 LN5-12/40: 2D+Color, Triple, CPA & Triple CPAIndex Label M.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index

Strona 370 - Reference Manual 62

Reference Manual 123 LN5-12LN5-12: 2D & 2D+M modeIndex Label M.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index Value 1.500 0.70

Strona 371 - Reference Manual 63

Reference Manual 124 LN5-12: Pulsed Doppler & 2D+Pulsed DopplerIndex Label M.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index Val

Strona 372 - Reference Manual 64

Reference Manual 125 EC4-9: 2D+Color, Triple, CPA & Triple CPAIndex LabelM.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index Value

Strona 373 - Reference Manual 65

Reference Manual 126 EVN4-9EVN4-9: 2D & 2D+M modeIndex LabelM.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index Value 1.360 0.581

Strona 374 - Reference Manual 66

Reference Manual 127 EVN4-9: Pulsed Doppler & 2D+Pulsed DopplerIndex LabelM.I.TIS TIBTICscannon-scannon-scanAaprt≤1 Aaprt>1Maximum Index Valu

Strona 375 - Reference Manual 67

Reference Manual 128 Wo1x1 : For non–scanning modes and/or non–scanning components of combi-national modes: time average acoustic power at the s

Strona 376 - Reference Manual 68

Reference Manual 129 C2-4/20C2-4/20: 2D & 2D+M modeIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²Maximum index

Strona 377 - Reference Manual 69

Reference Manual 130 C2-4/20: Pulsed Doppler & 2D+Pulsed DopplerIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²Ma

Strona 378 - Reference Manual 70

Reference Manual 131 C2-5: 2D+Color, Triple, CPA & Triple CPAIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²Maxim

Strona 379 - Reference Manual 71

1-28User Manual M-mode and Doppler Controls SpeedM-mode and Doppler sweep speed adjustments will not aect the MI. When M-mode sweep speed changes,

Strona 380 - Reference Manual 72

Reference Manual 132 C2-8C2-8: 2D & 2D+M modeIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²Maximum index value

Strona 381 - Reference Manual 73

Reference Manual 133 C2-8: Pulsed Doppler & 2D+Pulsed DopplerIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²Maxim

Strona 382 - Reference Manual 74

Reference Manual 134 CN2-8: 2D+Color, Triple, CPA & Triple CPAIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²Maxi

Strona 383 - Reference Manual 75

Reference Manual 135 CN4-9CN4-9: 2D & 2D+M modeIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²Maximum index val

Strona 384 - Reference Manual 76

Reference Manual 136 CN4-9: Pulsed Doppler & 2D+Pulsed DopplerIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²Maxi

Strona 385 - Reference Manual 77

Reference Manual 137 L5-12/60: 2D+Color, Triple, CPA & Triple CPAIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²M

Strona 386 - Reference Manual 78

Reference Manual 138 LE5-12LE5-12: 2D & 2D+M modeIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²Maximum index va

Strona 387 - Reference Manual 79

Reference Manual 139 LE5-12: Pulsed Doppler & 2D+Pulsed DopplerIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²Max

Strona 388 - Reference Manual 80

Reference Manual 140 LN5-12/40: 2D+Color, Triple, CPA & Triple CPAIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²

Strona 389 - Reference Manual 81

Reference Manual 141 LN5-12LN5-12: 2D & 2D+M modeIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²Maximum index va

Strona 390 - Reference Manual 82

1-29Chapter 1 Safety  ApplicationAcoustic output defaults are set when you select an application. Samsung Medison factory defaults vary with probe,

Strona 391 - Reference Manual 83

Reference Manual 142 LN5-12: Pulsed Doppler & 2D+Pulsed DopplerIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²Max

Strona 392 - Reference Manual 84

Reference Manual 143 EC4-9: 2D+Color, Triple, CPA & Triple CPAIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²Maxi

Strona 393 - Reference Manual 85

Reference Manual 144 EVN4-9EVN4-9: 2D & 2D+M modeIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²Maximum index va

Strona 394 - Reference Manual 86

Reference Manual 145 EVN4-9: Pulsed Doppler & 2D+Pulsed DopplerIndex Label MI ScanTIS TIBTICNon-scan Non-scanNon-scanAaprt≤1cm² Aaprt>1cm²Max

Strona 396 - Reference Manual 88

1-30User ManualThe acoustic output for this system has been measured and calculated in accordance with the December 1985 “510(K) Guide for Measuring a

Strona 397 - Reference Manual 89

1-31Chapter 1 Safety Acoustic Output and MeasurementThe terms and symbols used in the acoustic output tables are dened in the following paragraph

Strona 398 - Reference Manual 90

1-32User Manual Acoustic Measurement Precision and UncertaintyThe Acoustic Measurement Precision and Acoustic Measurement Uncertainty are described b

Strona 399 - Reference Manual 91

1-33Chapter 1 Safety The document includes analysis and discussion of the sources of error / measurement eects due to:XBalance system calibration.

Strona 400 - Reference Manual 92

SAMSUNG MEDISON DIAGNOSTIC ULTRASOUND SYSTEMVersion 1.02EnglishSONOACE R5User Manual

Strona 401 - Reference Manual 93

1-34User Manual Environmental ProtectionCAUTION:XThe console and peripherals could be sent back to manufacturers for recycling or proper disposal af

Strona 402 - Reference Manual 94

Chapter 2IntroductionSpecications ... 2-3Product Conguration ... 2

Strona 404 - 2D Measure

2-3Chapter 2 Introduction SpecicationsPhysical DimensionsHeight: 1,370mm (with handle)Width: 450mm (with probe holder)Depth: 700mm (with control

Strona 405 - & Febiger

2-4User ManualApplicationObstetrics, Fetal Heart, Abdomen, Gynecology, Renal, Urology, Breast, Vascular, Small Part, Musculoskeletal, Neonatal, Pediat

Strona 406 - Cardiology M mode

2-5Chapter 2 Introduction Product CongurationThe SONOACE R5 consists of the monitor, the control panel, the console, the peripheral devices, and

Strona 407 - Reference Manual 99

2-6User Manual Title AreaThis section displays the Logo, Patient Name, Hospital Name, Application, Frame Rate & Depth, Probe Information and Date

Strona 408 - Reference Manual 100

2-7Chapter 2 Introduction Control PanelThe control panel can be used for controlling the system. [Figure 2.2 Control Panel]The user can manipulate t

Strona 409 - Cardiology C mode

2-8User ManualOn/OTurns the system on/o.PatientDisplays the Patient Information screen for patient selection and information entry.End ExamFinishes

Strona 410 - Cardiology Doppler

2-9Chapter 2 Introduction BodyMarkerThis is used to start or nish the BodyMarker Mode.IndicatorThis is used to start or nish the indicator Mode.Tr

Strona 412 - Vascular Reference

2-10User ManualCAUTION: Too large a difference in the gain value settings of adjacent TGC slides may lead to the generation of stripes in an image. Sof

Strona 413 - Urology Reference

2-11Chapter 2 Introduction F3BiopsyThis is used to start the biopsy.F4Data on / oThis button is used to display information of image on the screen

Strona 414 - Reference Manual 106

2-12User ManualConsoleWhilst the inside of the console consists of devices enabling the implementation of ultrasound images, the outside of the consol

Strona 415 - Fetal Echo Reference

2-13Chapter 2 Introduction Rear PanelA monitor, printer, VCR, etc. are connected via the rear panel at the back of the system. [Figure 2.6 Rear Pa

Strona 416 - Preload Index

2-14User Manual Power Connection PartThe power connection part is located at the back of the system.162354[Figure 2.7 Power Connection Part]Item Desc

Strona 417 - Acoustic Power Tables

2-15Chapter 2 Introduction NOTE: Cover the power inlet area with a plug bracket and secure it with screws. Probe HolderA probe holder is mounted at

Strona 418 - Reference Manual 110

2-16User ManualThe following products are recommended:  DVDDVD-RW, DVD-RAM, DVD-R, DVD+R, DVD-ROM, CD-R, CD-RW, CD-ROM Video Page printer − Black a

Strona 419 - Reference Manual 111

2-17Chapter 2 Introduction  OthersFlash Memory mediaNOTE:XIf you use the USB 1.1 flash memory, the system cannot recognize it. In the case of this

Strona 420 - Reference Manual 112

2-18User ManualProbeProbes are devices that generate ultrasound waves and process reected wave data for the purpose of image formation.NOTE: For more

Strona 421 - Reference Manual 113

2-19Chapter 2 Introduction Accessory This system provides a box including accessories.CAUTION: Main cord set, separately certified according to the r

Strona 422 - Reference Manual 114

PROPRIETRAY INFORMATION AND SOFTWARE LICENSEThe Customer shall keep condential all proprietary information furnished or disclosed to the Customer by

Strona 423 - Reference Manual 115

2-20User ManualOptionsThe SONOACE R5 has the following optional functions:XImage ling-SONOVIEW XPulse wave DopplerXDICOMXColor DopplerXFreehand

Strona 424 - Reference Manual 116

Chapter 3Starting DiagnosisPower Supply ... 3-3Powering On ...

Strona 426 - Reference Manual 118

3-3Chapter 3 Starting Diagnosis Power SupplyBoot the system for use.CAUTION: Make sure to connect the probe and peripheral devices that will be u

Strona 427 - Reference Manual 119

3-4User Manual Probes & ApplicationsBefore scanning, select a probe and an application. CAUTION: Please refer to “Chapter 9. Probes” for more inf

Strona 428 - Reference Manual 120

3-5Chapter 3 Starting Diagnosis Probe and Application SelectionSet a probe and an application in the Applications screen.1. Select a probe and an

Strona 429 - Reference Manual 121

3-6User ManualEditing BodyMarker Select some BodyMarkers to use in the system or change the order of BodyMarkers. Click Edit BodyMarker in the Applica

Strona 430 - Reference Manual 122

3-7Chapter 3 Starting Diagnosis Add BodyMarkers Select a BodyMarker on ‘Group’ using the trackball. Press Set button to add to the ‘User dened Bo

Strona 431 - Reference Manual 123

3-8User ManualCalc Sequence[Figure 3.3 Calc Package Sequence Change]To change the order and content of the Calculation Package of each application, cl

Strona 432 - Reference Manual 124

3-9Chapter 3 Starting Diagnosis Patient InformationPress the Patient button on the control panel and then the Patient Information screen will be

Strona 433 - Reference Manual 125

Safety Requirements Classications: XType of protection against electrical shock: Class I XDegree of protection against electrical shock (Patient

Strona 434 - Reference Manual 126

3-10User ManualRegistering a New Patient1. After clicking the New in the Patient Information screen, enter the patient name, ID and other information

Strona 435 - FDA Tables

3-11Chapter 3 Starting Diagnosis  GAIt indicates the gestational age of a patient. You can enter it manually in the specied format, or have it au

Strona 436 - Reference Manual 128

3-12User Manual BSA If height and weight are entered, the body surface area (BSA) is automatically calculated and shown. HR Enter the patient’s hear

Strona 437 - Reference Manual 129

3-13Chapter 3 Starting Diagnosis Searching for Patient InformationYou can search the patient information with following three methods. Finding Pat

Strona 438 - Reference Manual 130

3-14User Manual Finding Patient Information using WorklistThe system connects to the hospital’s DICOM Modality Worklist server and obtains information

Strona 439 - Reference Manual 131

3-15Chapter 3 Starting Diagnosis Modifying Patient InformationTo modify information in the Patient Information screen:1. Modify the information us

Strona 441 - Reference Manual 133

Chapter 4Diagnosis ModesInformation ... 4-3Diagnosis Mode Types ...

Strona 443 - Reference Manual 135

4-3Chapter 4 Diagnosis Modes InformationDiagnosis Mode TypesThis product supports a variety of diagnosis modes including Basic Mode, Combined Mode,

Strona 444 - Reference Manual 136

XMedical Electrical Equipment, Part 2-37: Particular Requirements for Safety - Ultrasonic Medical Diagnostic and Monitoring Equipment [IEC 60601-2-37

Strona 445 - Reference Manual 137

4-4User ManualBasic UseThe items that can be used commonly in each diagnosis mode are shown below: QuickScan / Gain XQuickScan: The ‘Q Scan’ mark wi

Strona 446 - Reference Manual 138

4-5Chapter 4 Diagnosis Modes ZoomPress the Zoom button on the control panel to start Read Zoom mode. This function allows you to zoom an image saved

Strona 447 - Reference Manual 139

4-6User Manual Harmonic Press the HAR button on the control panel to select On or O. ‘HGen’, ‘HRes’. ‘HPen’ mark is displayed in the image informati

Strona 448 - Reference Manual 140

4-7Chapter 4 Diagnosis Modes Basic Modes2D ModeThis basic mode, also referred to as B Mode (Brightness mode), provides scan planes of organs. This i

Strona 449 - Reference Manual 141

4-8User Manual Starting 2D ModeNOTE: As 2D Mode is included in all diagnosis modes, it cannot be closed. Press the 2D button on the control panel. Pre

Strona 450 - Reference Manual 142

4-9Chapter 4 Diagnosis Modes FSI FSI stands for Full Spectrum Imaging. Select a value between 1 - 3 with the Menu dial-button.This function synthesi

Strona 451 - Reference Manual 143

4-10User Manual Trapezoid NOTE: This item appears in the Softmenu only when a Linear Probe is used.In general, the rectangular frame provided by a Li

Strona 452 - Reference Manual 144

4-11Chapter 4 Diagnosis ModesTips!Editing a post mapYou can edit the user post map with following two methods. 1. Select the each point of the map w

Strona 453 - Reference Manual 145

4-12User Manual 2D Mode SoftmenuFollowing items are displayed on Softmenu in 2D mode. Select the item and adjust the value with the Softmenu button [1

Strona 454

4-13Chapter 4 Diagnosis ModesM ModeBy drawing the M line across a desired area within the 2D image, you can view changes over time.This is useful for