Chapter 5 Measurements and Calculations
Pediatric Hips Calculation
Measurement Menu
Measured items are automatically recorded in a report.
[Figure 5.19 Pediatric Hips Measurement Menu]
Measurement Menu Item Method Unit
Hip Angle Type Three Distance Measurement °
Measurement Method
1.
Use the
Trackball
and the
Set
button on the control panel to specify the rst straight line.
X Place the cursor at a desired position with the Trackball, and press the Set button.
Tips!
Repositioning Point
Pressing the Change button before pressing the Set button to complete positioning resets the
position of a point just set.
2.
Repeat the above process to specify two other straight lines.
3.
The angle between them will be calculated automatically.
X α: The angle between the rst and second straight lines.
X β: The angle between the rst and third straight lines.
4.
When the measurement is nished, its result is shown on the screen.