5 - 8 SONOACE X6 Operation manual
(2) Distance is displayed with cmand with the second digit.
M Mode
Measurements:
System Tolerance:
(whichever is greater)
Test
Methodology
Accuracy
(1)
Based on
Range
(2)
Depth :
<+/- 5% or 3 mm
Phantom
Phantom
1 - 25 cm
Time:
< +/- 5%.
Signal generator
Phantom
.01 - 11.3 sec
(1) Measurement accuracy depends on the table and human errors.
(2) Distance is displayed with cmand with the second digit. Velocity is displayed with cm/sand with
the second digit.
Spectral Doppler Mode
Doppler
Measurement
System Tolerance
(whichever is greater)
Methodology Range
(2)
Velocity < ± 15% Phantom
PW:0.1cm/s 8.8 m/s
CW:.1cm/s - 19.3 m/s
Time:
< +/- 5%.
Signal generator
.01 - 11.3 sec
(1) Measurement accuracy depends on the table and human errors.
(2) Distance is displayed with cmand with the second digit. Velocity is displayed with cm/sand with
the second digit.