Safety Requirements
Classication:
Type of protection against electrical shock: Class I
Degree of protection against electrical shock (Patient connection): Type BF applied part
Degree of protection against harmful ingress of water: Ordinary equipment
Degree of safety of application in the presence of a ammable anesthetic material with air or with
oxygen or nitrous oxide: Equipment not suitable for use in the presence of a ammable anesthetic
mixture with air or with oxygen or nitrous oxide.
Mode of operation: Continuous operation
Electromechanical safety standards met:
Medical Electrical Equipment, Part 1: General Requirements for Basic Safety and Essential
Performance [IEC 60601-1: 2005]
Medical Electrical Equipment, Part 1-2: General Requirements for Basic Safety and Essential
Performance- Collateral Standard: Electromagnetic Compatibility - Requirements and Tests [IEC
60601-1-2: 2007]
Medical Electrical Equipment, Part 1-6: General Requirements for Basic Safety and Essential
Performance- Collateral Standard: Usability [IEC 60601-1-6: 2010]
Medical Electrical Equipment, Part 2-37: Particular Requirements for the Basic Safety and Essential
Performance of Ultrasonic Medical Diagnostic and Monitoring Equipment [IEC 60601-2-37: 2007]
Medical Electrical Equipment, Part 1: General Requirements for Safety [IEC 60601-1: 1988 with A1:
1991 and A2: 1995]
Medical Electrical Equipment, Part 1-1: General Requirements for Safety - Collateral Standard: safety
Requirement for Medical Electrical Systems [IEC 60601-1-1: 2000]
Medical Electrical Equipment, Part 1-2: General Requirements for Safety - Collateral Standard:
Electromagnetic Compatibility - Requirements and Test [IEC 60601-1-2: 2001, A1: 2004]
Medical Electrical Equipment, Part 1-4: General Requirements for Safety - Collateral Standard:
Programmable Electrical Medical Systems [IEC 60601-1-4: 1996, A1: 1999]