Operation Manual
It is not recommended to use images in Color/Power Doppler Mode for measurement of accurate
blood ow velocity.
The amount of blood ow is calculated based on the average velocity rather than the peak velocity.
In all applications, the amount of blood ow is measured in PW/CW Spectral Doppler Mode.
Cursor Position
All measurements are aected by input data.
To ensure accurate positioning of the cursor:
Adjust the images on the screen so that they are displayed at maximum granularity.
Use the front edge or boundary point of a probe to make the start and end points of a measurement
object more distinct.
Make sure that the probe direction is always aligned during measurement.