GE HEALTHCARE
D
IRECTION 5194296-100, REVISION 10 LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
Chapter 4 Functional Checks 4-23
4-3-11-3 Worksheets
Measurement/Calculation worksheets are available to display and edit measurements and calculations.
There are generic worksheets as well as Application specific worksheets. The worksheets are selected
from the Measurement Touch Panel.
4-3-11-4 Report* Pages
Measurements/Calculations that are included on the worksheet can also be displayed on Report Pages.
Report Pages can be customized to meet the appropriate needs of the user.
4-3-12 Probe/Connectors Usage
4-3-12-1 Connecting a probe
1.) Place the probe's carrying case on a stable surface and open the case.
2.) Carefully remove the probe and unwrap the probe cable.
3.) DO NOT allow the probe head to hang free. Impact to the probe head could result in irreparable
damage.
4.) Align the connector with the probe port and carefully push into place.
5.) Lock the probe latch upward.
6.) Carefully position the probe cord so it is free to move and is not resting on the floor.
4-3-12-2 Activating the probe
The probe activates in the currently-selected operating mode. The probe's default settings for the mode
and selected exam are used automatically.
4-3-12-3 Deactivating the probe
When deactivating the probe, the probe is automatically placed in standby mode.
1.) Press the Freeze key.
2.) Gently wipe the excess gel from the face of the probe. (Refer to the Basic User Manual for
complete probe cleaning instructions.)
3.) Carefully slide the probe around the right side of the keyboard, toward the probe holder.
Ensure that the probe is placed gently in the probe holder.
4-3-12-4 Disconnecting the probe
Probes can be disconnected at any time. However, the probe should not be selected as the active
probe.
1.) Unlock the probe latch downward.
2.) Pull the probe and connector straight out of the probe port.
3.) Carefully slide the probe and connector away from the probe port and around the right side of the
keyboard.
4.) Ensure the cable is free.
5.) Be sure that the probe head is clean before placing the probe in its storage box.
NOTICE
* Report is not available on LOGIQ Book XP PRO.
WARNINGWARNING
Take the following precautions with the probe cables: Do not bend. If you have
purchased the cart option, be sure to keep probe cables free from the wheels.
WARNINGWARNING
Be careful not to trip on the probe cables if using the device without the optional cart.