Samsung SV-651B Instrukcja Obsługi Strona 126

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 125
GE HEALTHCARE
DIRECTION 5194296-100, REVISION 10 LOGIQ BOOK XP SERIES BASIC SERVICE MANUAL
4-20 Section 4-3 - General Procedure
4-3-10-2 Doppler Mode OP Panel Controls
Table 4-9 Doppler Mode OP Panel Controls
Step Task Expected Result(s)
1
Press PW Mode key PW Mode Starts
2
Adjust Gain
Amplifies the overall strength of the echoes processed in the Color Flow window.
Turn the Gain dial (PW Mode key) to the left/right to increase/decrease Gain.
3
Display M/D-Mode Cursor
Displays the M/D-Mode cursor on the B-Mode image. Press Cursor and Trackball to
position sample volume graphic. Click SV gate to adjust sample volume gate size.
4
B Pause
Toggle between simultaneous and update presentation while viewing Spectral
Doppler. Press B Pause to toggle between simultaneous and update.
Przeglądanie stron 125
1 2 ... 121 122 123 124 125 126 127 128 129 130 131 ... 225 226

Komentarze do niniejszej Instrukcji

Brak uwag