Operation Manual
5-12
2D Hip Joint
This is a basic measurement that is available in all diagnosis modes. You can measure the angle of two
straight lines to a straight line in a 2D image. This measurement is commonly used in the obstetrics appli-
cation.
1. Press the Softmenu button [4] 2D Hip Joint. “2D Hip Joint” is displayed on the left side of the screen.
2. Use the trackball and the Set button on the control panel to specify the rst straight line.
Place the cursor at a desired position with the trackball, and press the Set button.
3. Repeat the above process to specify two other straight lines.
4. The angle between them will be calculated automatically.
α: The angle between the rst and second straight lines.
β: The angle between the rst and third straight lines.
5. When the measurement is nished, its result is shown on the screen. The angles α and β, and the Hip
Joint Type are displayed on the screen.
Please refer to the table below for Hip Joint Type information:
Type α β
1a 60 ≤ α
90
0
β
55
1b 60 ≤ α
90
55 ≤ β
90
2a/b 50 ≤ α
60
0
β
90
2c 43 ≤ α
50
77 ≤ β
90
d 43 ≤ α
50
0
β
77
3/4 0
α
43
[Table 5.2 Hip Joint Type Table]
M Distance
This is a basic measurement that is available in M Mode only. You can specify two points in an M image
and measure the distance, elapsed time and velocity between them.
1. Press the Softmenu button [1] Distance to select M Distance. “M Distance is displayed on the left side
of the screen.
2. Specify two points and measure the straight distance between them. The method for measuring a
line is the same as in “2D Distance (Line).
3. When the measurement is nished, its result is shown on the screen.