Chapter 9 Probes
9-11
Use and Infection Control of the Probe
WARNING: No neurosurgical treatments or examinations should be carried out on a patient
with Creutzfeldt-Jakob disease (critical brain disease caused by virus). If the probe has been
used on such a patient, it cannot be sterilized by any method whatsoever.
CAUTION: Sufficient washing and disinfecting must be carried out for preventing infection.
This is the responsibility of the user who manages and maintains the disinfection procedures
for the equipment. Always use legally approved detergents and sheaths.
The ultrasonographic image scanner uses ultrasound, and it makes direct contact with the patient when
in use. Depending on the types of examinations, such contact can be made to a wide variety of locations
including the ordinary skin or the location of blood transfusion during a surgery.
The most eective method to prevent infection among patients is to use each probe only once. However,
probes may need to be re-used as they are complex in design and expensive. Consequently, protective
devices such as sheaths must be used, and the safety instructions must be followed carefully in order to
minimize the risk of infection among patients.
Electric Shocks
The probe uses electrical energy. If it touches conductive materials, there are risks of electric shocks to the
patient or the user.
WARNING:
X
Regularly receive short-circuit examination from the Samsung Medison Customer Support
Department.Do not immerse the probe into liquid.
X
Do not immerse the probe into liquid
X
Do not drop the probe or apply mechanical shocks.
X
Inspect the housing, strain relief, lens and seal for damage, and check for any functional problem
before and after each use.
X
Do not apply excessive force to twist, pull or bend the probe cable. It may result in a short circuit.
X
The power protection fuse protects the probe and the product from excess current. If the power
monitoring protection circuit detects excess current, it immediately shuts off the current to the
probe in order to prevent the probe surface from overheating and to restrict the ultrasound
power output.
The temperature of the product for making contact with patients is limited under 43°C. The ultrasound
power output (AP&I) is in compliance with US FDA standards.