Chapter 1 Safety
1-7
Electrical Safety
This equipment has been veried as a Class I device with Type BF applied parts.
CAUTION: As for US requirement, the LEAKAGE CURRENT might be measured from a center-tapped
circuit when the equipment connects in the United States to 240V supply system.
Prevention of Electric Shock
A
dditional equipment connected to medical electrical equipment must comply with the respective IEC or ISO
standards (e.g. IEC 60950 for data processing equipment). Furthermore all configurations shall comply with
the requirements for medical electrical systems (see IEC 60601-1-1 or clause 16 of the 3 Ed. of IEC 60601-1,
respectively). Anybody connecting additional equipment to medical electrical equipment congures a medical
system and is therefore responsible that the system complies with the requirements for medical electrical systems.
Attention is drawn to the fact that local laws take priority over the above-mentioned requirements. If in doubt,
consult your local representative or the technical service department.
WARNING:
X
Electric shock may exist result if this system, including and all of its externally mounted recording
and monitoring devices, is not properly grounded.
X
Do not remove the covers on the system; hazardous voltages are present inside. Cabinet panels
must be in place while the system is in use. All internal adjustments and replacements must be
made by a qualified Samsung Medison Customer Service Department.
X
Check the face, housing, and cable before use. Do not use and disconnect the power source, if
the face is cracked, chipped, or torn, the housing is damaged, or if the cable is abraded.
X
Always disconnect the system from the wall outlet prior to cleaning the system.
X
All patient contact devices, such as probes and ECG leads, must be removed from the patient
prior to application of a high voltage defibrillation pulse.
X
Never use the product in proximity to any flammable anesthetic gases (N
2
O) or oxidizing gases.
There is a risk of explosion.
X
Avoid places where the system is likely to be difficult to operate the disconnection device.