EMC
The testing for EMC(ElectroMagnetic Compatibility) of this system has been performed according to the
international standard for EMC with medical devices (IEC 60601-1-2). This IEC standard was adopted in Europe
as the European norm (EN 60601-1-2).
Guidance and manufacturer’s declaration - electromagnetic emission
This product is intended for use in the electromagnetic environment specied below. The customer or the
user of this product should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment -guidance
RF Emission
CISPR 11
Group 1
The Ultrasound System uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF Emission
CISPR 11
Class A
The Ultrasound System is suitable for use in all establishments
other than domestic, and may be used in domestic establishments
and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes, provided the following warning is heeded:
Warning: This system is intended for use by healthcare
professionals only. This system may cause radio interference or may
disrupt the operation of nearby equipment. It may be necessary
to take mitigation measures, such as re-orienting or relocating the
Ultrasound System or shielding the location.
Harmonic Emission
IEC 61000-3-2
Class A
Flicker Emission
IEC 61000-3-3
Complies