Samsung SONOACE R3 Instrukcja Użytkownika Strona 41

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 442
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 40
Chapter 1 Safety
1-23
The measured water tank values are de-rated using a conservative, industry standard, attenuation
coecient of 0.3dB/cm-MHz.
Conservative values for tissue characteristics were selected for use in the TI models. Conservative values
for tissue or bone absorption rates, blood perfusion rates, blood heat capacity, and tissue thermal
conductivity were selected.
Steady state temperature rise is assumed in the industry standard TI models, and the assumption is
made that the ultrasound probe is held steady in one position long enough for steady state to be
reached.
A number of factors are considered when estimating the accuracy of display values: hardware variations,
algorithm accuracy estimation and measurement variability. Variability among probes and systems
is a significant factor. Probe variability results from piezoelectric crystal efficiencies, process-related
impedance dierences, and sensitive lens focusing parameter variations. Dierences in the system pulse
voltage control and eciencies are also a contributor to variability. There are inherent uncertainties in
the algorithms used for estimating acoustic output values over the range of possible system operating
conditions and pulse voltages. Inaccuracies in laboratory measurements are related to differences
in hydrophone calibration and performance, positioning, alignment and digitization tolerances, and
variability among test operators.
The conservative assumptions of the output estimation algorithms of linear propagation, at all depths,
through a 0.3dB/cm-MHz attenuated medium are not taken into account in calculation of the accuracy
estimate displayed. Neither linear propagation, nor uniform attenuation at the 0.3dB/cm-MHz rate,
occur in water tank measurements or in most tissue paths in the body. In the body, dierent tissues and
organs have dissimilar attenuation characteristics. In water, there is almost no attenuation. In the body,
and particularly in water tank measurements, non-linear propagation and saturation losses occur as
pulse voltages increase.
The display accuracy estimates take into account the variability ranges of probes and systems, inherent
acoustic output modeling errors, and measurement variability. Display accuracy estimates are not based
on errors in, or caused by measuring according to, the AIUM measurement standards. They are also
independent of the eects of non-linear loss on the measured values.
Control aecting the indices
As various system controls are adjusted, the TI and MI values may change. This will be most apparent as
the POWER control is adjusted; however, other system controls will aect the on-screen output values.
Power
Power controls the system acoustic output. Two real-time output values are on the screen: a TI and a MI.
They change as the system responds to POWER adjustments.
In combined modes, such as simultaneous Color, 2D-mode and pulsed Doppler, the individual modes
each add to the total TI. One mode will be the dominant contributor to this total. The displayed MI will
be from the mode with the largest peak pressure.
Przeglądanie stron 40
1 2 ... 36 37 38 39 40 41 42 43 44 45 46 ... 441 442

Komentarze do niniejszej Instrukcji

Brak uwag